LDM5 Meter Setup and Operation Instructions
12 312668W
Calibration (FIG. 13)
Recalibrates the meter for dispensing different fluids.
1. If the Calibration Menu is not already displayed,
complete steps 1 - 4 of the Setup Menus section,
page 10.
2. CAL blinks indicating you are on the Calibration
Menu screen. Do ONE of the following.
Use the current calibration.
• Press and hold Reset but-
ton to lock in amount. Dis-
play returns to the
Standard Mode.
OR
Recalibrate meter as follows:
a. If the unit of measurement is gallons, pints,
or quarts, dispense exactly one quart of fluid
into a calibrated 1–quart container.
If the unit of measurement is liters, dispense
exactly one liter of fluid into a calibrated
1–liter container.
For proper calibration, you must dis-
pense the exact amount.
b. Press and hold Total
button until CAL stops
blinking and the dis-
play shows 1.00.
c. When CAL starts to
blink again, the display
should show 1.00. The
new calibration is complete.
3. Press and hold the Reset but-
ton. The unit returns to Stan-
dard Dispense Mode.
F
IG. 13
GAL
QTSPTS
LITERS
R-TOTAL
CAL
Min. Dispense Volume 0.5 liter (0.13 gal)
Minimum Flow Rate 1.0 l/min. (0.26 gpm)
Maximum Flow Rate 19.0 l/min. (5.00 gpm)
Maximum Working Pressure: 70 bar (1000 PSI)
If an error was made during meter recalibration,
repeat Steps a - c of the recalibration process to
recalibrate the meter again.